Interpreting "Rust" - Yellow-brown to Dark-brown Stains or Spots
Yellow-brown to dark-brown stains or spots on surgical stainless steel instruments are frequently mistaken for "rust". These residue deposits (stains or spots arranged in groups or along edges or in crevices) are usually the instrument being exposed to result of high chloride content. They will lead to pitting of the surgical instrument surface if not removed. (see Avoiding High Levels of Chloride below) Excessively hard water can contain high levels of salt sufficient to cause stains or spots that appear as rust. Boilers used to generate the steam for steam sterilizers, if not cleaned properly, will produce contaminated steam which can deposit minerals onto instruments during the sterilization process.
ALSO AVOID using cleaning agents containing other halogen ion agents
Use cleaning agents containing "Nonionic Surfactants" whenever possible.
Virtually all manufacturers of surgical instruments, rigid scopes, flexible scopes, and instrument containers recommend the use of neutral pH Cleaning Concentrates. Generic Example of this recommendation: Do not use high acidic (pH <4) or high alkaline (pH >10) products for disinfection or cleaning, since these can corrode metal, cause discoloration or stress fractures.
Do not use abrasive pads or abrasive Cleaning Concentrates , which will scratch the surface allowing dirt and water deposits to collect. Abrasive cleaning will remove the protective passive layer.
Do not use Cleaning Concentrates with high concentrations of chlorine bleach to clean or disinfect stainless steel instruments, as pitting will occur. Never use bleach to clean any surgical instruments. The high pH of bleach causes surface deposits of brown stains and might even corrode the instrument. Even high quality stainless steel is not impervious to an acidic bleach solution.
Sort instruments by similar metal for subsequent processing so that electrolytic deposition (galvanic corrosion) due to contact between dissimilar metals will not occur.
(Guidelines on metals and alloys in contact with food; Council of Europe; published 11.10.2000. Systemic nickel: the contribution made by stainless steel cooking utensils; Contact Dermatitis, Volume 32:2, 1994) of the stainless steel passive layer to prevent corrosion have revealed a reduction in corrosion prevention with the use of cleaning concentrates that are not neutral pH. The use of cleaning concentrates that deliver an acid rinse will release nickel from the stainless steel and decrease the efficacy of the passive layer. This is most critical on initial reprocessing events of stainless steel surgical instruments. Measurable levels of nickel have been detected. It was also shown that, as the number of subsequent uses increased, the level of nickel release diminished and reached a steady state (measured in the order of μg/l). These observations reflect the changes that occur in the passive oxide layer on first immersion of stainless steels in aqueous media.
Stainless steel is essentially a low carbon steel which contains chromium at 10% or more by weight. It is this addition of chromium that gives the steel its unique stainless, corrosion resisting properties. The chromium content of the steel allows the formation of a rough, adherent, invisible, corrosion-resisting chromium oxide film on the steel surface. If damaged mechanically or chemically, this film is self-healing, providing that oxygen, even in very small amounts, is present. The corrosion resistance and other useful properties of the steel are enhanced by increased chromium content and the addition of other elements such as molybdenum, nickel and nitrogen. Stainless steel has a passive film created by the presence of chromium (and often other alloying elements, nickel, molybdenum) that resists this process. When exposed in air, stainless steels passivate naturally (due to the presence of chromium). But the time required can vary. In order to ensure that the passive layer reforms rapidly after pickling, a passivation treatment is performed using a solution of nitric acid and water.
How is the "passive oxide layer" Manufactured and Maintained ?
The passive layer or stainless steel is intended to prevent or resist corrosion. The process is called “Passivation”. “Passivation” and Polishing eliminate the carbon molecules form the instrument surface. This forms a layer which acts as a corrosive resistant seal. Passivation is a chemical process that removes carbon molecules from the surface of the instrument. This chemical process can also occur through repeated exposure to oxidizing agents in chemicals, soaps, and the atmosphere. Polishing, by the manufacturer, is a process used to achieve a smooth surface on the instrument. Surgical Instruments are polished because the passivation process leaves microscopic pits where the carbon molecules were removed. Polishing also builds a layer of chromium oxide on the surface of the surgery instrument. Proper cleaning, handling, and sterilization will build up the layer of chromium oxide and protect the Surgical Instrument from corrosion and /or pitting. In some circumstances older instruments have higher resistance to corrosion than new ones. The newer instruments have not had the time to build up the chromium oxide layer. Improper cleaning and sterilization can cause the layer of chromium oxide to disappear or become damaged thus increasing the possibility of corrosion and/or pitting. Proper cleaning and sterilization can cause the layer of chromium oxide to improve over time thus decreasing the possibility of corrosion and/or pitting. Second only to the financial asset value of the working staff, the surgery instrument and scope inventory is the single most financially valuable asset of the healthcare facility. It is important to properly clean, sterilize, handle, and store your instruments.
John Temple Product Development


Pre wash and soaking Surgical Instruments
Surgical Instrument Decontamination
Automated Surgical Instrument Washers
Ultrasonic Surgical Instrument Cleaners
Pre wash and soaking Surgical Instruments
Surgical Instrument Decontamination
Automated Surgical Instrument Washers
Ultrasonic Surgical Instrument Cleaners